Pendulum Impact Tester (Charpy/Izod) | 556-IT
Pendulum Impact Tester (Charpy/Izod) | 556-IT
ISO 179, ISO 180, ISO 8256, ASTM D256, ASTM D1822
The (Pendulum) Impact Tester measures the impact value from the amount of pendulum energy required to fracture a specimen.
The Izod, Charpy, and Tensile methods can be used.
Optional instrumentation is useful for analysis of brittleness and toughness at impact.
| Test methods | Specimen support/fixture methods | Testing standards | |
| Charpy impact test | Fix both ends of the specimen to a horizontal support and impact the back of the notch (notch) with a hammer. | ![]() |
• ISO 179 (JIS K 7111)
• ASTM D6110 |
| Izod impact test | Fix the end of the specimen in a vertical cantilevered beam and impact it with a hammer from the notched direction. | ![]() |
• ISO 180 (JIS K 7110)
• ASTM D256 |
Related standards
| Charpy | • ISO 179 (JIS K 7111)
• ASTM D256 |
| Izod | • ISO 180 (JIS K 7110)
• ASTM D256 |
| Tensile-impact | • ISO 8256 (JIS K 7160)
• ASTM D1822 |
Features
– Digital display type eliminates abrasion loss error of the needle and reading error by the measurer.
– Charpy impact test, Izod impact test, tensile impact test, etc. can be performed by changing the hammer and specimen support.
– Easy operation with color LCD touch panel. Average, maximum, minimum, coefficient of variation, and standard deviation of impact values can be displayed and saved.
– Measurement data can be saved on a PC*. (*Optional dedicated data processing software is required.)
– Hammer release* is available at the touch of a button. No individual differences in hook release and improved reproducibility. (*Optional)
– Instrumentation system* selectable. (*Optional. Charpy impact test and in-base method tensile impact test only)
– Specimen support stand* with positioning can be selected. Specimens can be set easily and precisely. (*Option for Charpy impact test)
– Safety cover* (without or with top panel) selectable (*Optional)
Main unit
| Model | IT |
| Test methods | • Charpy impact test
• Izod impact test • Tensile impact test • Impact peeling test |
| Hammer capacity | • Charpy impact test: 0.5 to 15J
• Izod impact test: 0.5 to 22J • Tensile impact test: 1 to 15J |
| Hammer lifting angle | 150° |
| Power supply | Universal
(Single-phase, AC100-240V, 50/60Hz, 0.1kVA) |
| Dimensions | W600 × D360 × H1000mm |
| Weight (approx.) | 100kg |
Hammers
(1) Charpy impact test
ISO 179
| Hammer capacity | Model | Impact velocity | Impact blade tip |
| 0.5J / 1J* | JCH05J | 2.9m/sec. | R2mm、30° |
| 2J / 4J* | JCH2J | ||
| 5J | JCH5J | ||
| 7.5J / 15J* | JCH75J | 3.8m/sec. |
ASTM D6110
| Hammer capacity | Model | Impact velocity | Impact blade tip |
| 1J / 2J* | ACH1J | 3.46m/sec. | R3.17mm、45° |
| 3J / 6J* | ACH3J | ||
| 8J / 15J* | ACH8J |
(2) Izod impact test
ISO 180
| Hammer capacity | Model | Impact velocity | Impact blade tip |
| 0.5J / 1J* | JIZ05J | 3.5m/sec. | R0.8mm |
| 1J | JIZ1J | ||
| 2.75J / 5.5J* | JIZ27J | ||
| 11J / 22J* | JIZ11J |
ASTM D256
| Hammer capacity | Model | Impact velocity | Impact blade tip |
| 1J / 2J* | AIZ1J | 3.46m/sec. | R0.8mm |
| 3J / 6J* | AIZ3J | ||
| 8J / 15J* | AIZ8J |
*Additional weight
Specimen Support/Fixture
(1) Charpy impact test
JIS K 7111(ISO 179)
| Description | Model | Span |
| Charpy, ISO 179 specimen support (Standard) | JCHBAS | 62mm |
| Charpy, ISO 179 specimen support with centering unit (Optional) | JCBAST |
ASTM D6110
| Description | Model | Span |
| Charpy, ASTM D6110 specimen support | ACHBAS | 101.6mm |
(2) Izod impact test
ISO 180 / ASTM D256
| Description | Model | Distance between the top of the specimen fixture and the impact blade |
| Izod, specimen fixture | IZBASE | 22mm |
Options
| Hammer release device | ![]() ![]() |
| Curtain for litter control | ![]() |
| Safety cover | ![]() |
| Safety cover (Full cover type) | ![]() |
| Anchor weight | |
| Hammer brake | ![]() |
| Charpy specimen support with centering unit | ![]() |
| Small printer | |
| Data acquisition software |
Contact
-
Inquiries from the web
Contact Form -
Inquiries by telephone









Low Temperature Pendulum Impact Tester (Charpy/Izod) | 526-DG-40
Charpy/Izod Specimen Automatic Notching Machine / Notching Tool™ | 628-A-4/A-4E
Charpy/Izod Specimen Automatic Measuring Device | 138-NTM